TM 11-6625-3081-23
7-37. TEMPERATURE SENSORS
7-37
INPUT CODE:
AUGTEM (cont)
a .
DESCRIPTION (cont)
The A source is programmed to 30°C and the A auxiliary temperature sensor
checked after a 200-second delay, for an indication of 27.5° to 32.5°C.
If the indication is not within limits, a prompt is displayed.
The B source is programmed for 35°C.
The B controller temperature sensor is
checked after a 200-second delay, for an indication of 34.0° to 36.0°C.
If the indication is not within limits, a prompt is displayed.
The B source is programmed to 35°C.
The B auxiliary temperature is checked,
after a 200-second delay, for an indication of 32.5° to 37.5°C.
If the indication is not within the limits, a prompt is displayed.
The C source is programmed to 35°C and the C controller temperature sensor
checked, after a 200-second delay, for an indication of 34.0° to 36.0°C.
If the indication is not within limits, a prompt is displayed.
The C source is programmed to 35°C.
The C auxiliary temperature sensor checked,
after a 200-second delay, for an indication of 32.5° to 37.5°C.
If the indication is not within limits, a prompt is displayed.
7-189