TM 11-6625-3081-23
7-37. TEMPERATURE SENSORS
7-37
INPUT CODE: AU6TEM (cont)
k. SEGMENT 9
1. VDT DISPLAY
TEMP CONTROLLER SOURCE A IN ERROR TEMPERATURE READS #.## AND SHOULD READ
30.00.
PERFORM THE FOLLOWING REPLACE ACTIONS IN THE ORDER LISTED.
(REFERENCE TM 11-6625-3081-23 FOR REPLACEMENT PROCEDURES.)
REF. DES.
PART NO.
NAME
1. 1A1
13082765
CONTROL UNIT ASSY.
2. 2A3A1A1A2
13082971
SOURCE A
2. CORRECTIVE ACTION - SEGMENT 9
(a) Remove Control Unit Assembly 1A1.
Check contacts of mating connectors
and repair if required.
Install replacement 1A1 and perform AUGTEM
test.
(It is not mandatory to perform hand held temperature probe
portion of test.) If AUGTEM test passes, no further action required.
If this segment displayed again, proceed to (b). If another segment
displayed, proceed to that segment and continue troubleshooting.
(b) Remove Source A 2A3A1A1A2.
Check contacts of mating connectors and
repair if required.
Install replacement 2A3A1A1A2 and perform AUGTEM
test.
(It is not mandatory to perform hand held temperature probe
portion of test.) If AUGTEM test passes, no further action required.
If this segment displayed again, proceed to (c). If another segment
displayed, proceed to that segment and continue troubleshooting.
(c) Check the following cable assemblies and repair as required:
W1
1W4
2A3A1W6
If any cable was repaired, perform AUGTEM test. If AUGTEM test passes,
no further action required.
If AUGTEM test fails, repair is beyond
this level of maintenance.
Next higher level of maintenance required.
If no cables were repaired, repair is beyond this level of maintenance.
Next higher level of maintenance required.
If another segment displayed, proceed to that segment and continue
troubleshooting.