TM 11-6625-3081-23
7-37. TEMPERATURE SENSORS
7-37
INPUT CODE: AUGTEM (cont)
2. CORRECTIVE ACTION - SEGMENT 8 (cont)
(d)
(e)
(f)
(g)
Remove A/D Converter 2A2A13.
Check contacts of mating connectors and
repair if required.
Install replacement 2A2A13 and perform AUGTEM
test.
(It is not mandatory to perform hand held temperature portion of
test.) If AUGTEM test passes, no further action required. If this
segment displayed again, proceed to (e). If another segment displayed,
proceed to that segment and continue troubleshooting.
Remove Circuit and Heatsink 2A3A1A17XX (W23 SENSOR). Check contacts of
mating connectors and repair if required.
Install replacement
2A3A1A17XX and perform AUGTEM test.
(It is not mandatory to perform
hand held temperature probe portion of test.) If AUGTEM test passes,
no further action required.
If this segment displayed again, proceed
to (f).
If another segment displayed, proceed to that segment and
continue troubleshooting.
Remove Source C 2A3A1A17.
Check contacts of mating connectors and
repair if required.
Install replacement 2A3A1A17 and perform AUGTEM
test.
(It is not mandatory to perform hand held temperature portion of
test.) If AUGTEM test passes, no further action required. If this
segment displayed again, proceed to (g). If another segment displayed,
proceed to that segment and continue troubleshooting.
Check the following cable assemblies and repair as required:
W8
W11
1W16
2A3W3
2A3W4
2A3A1W5
If any cable was repaired, perform AUGTEM test. If AUGTEM test passes,
no further action required.
If AUGTEM test fails, repair is beyond
this level of maintenance.
Next higher level of maintenance required.
If no cables were repaired, repair is beyond this level of maintenance.
Next higher level of maintenance required.
If another segment displayed, proceed to that segment and continue
troubleshooting.
7-200