TM 11-6625-3081-23
7-37. TEMPERATURE SENSORS
7-37
INPUT CODE:
AUGTEM (cont)
2. CORRECTIVE ACTION - SEGMENT 10 (cont)
(d) Check the following cable assemblies and repair as required:
W11
1W16
2A3W3
2A3A1W5
If any cable was repaired, perform AUGTEM test. If AUGTEM test passes,
no further action required.
If AUGTEM test fails, repair is beyond
this level of maintenance.
Next higher level of maintenance required.
If no cables were repaired, repair is beyond this level of maintenance..
Next higher level of maintenance required.
If another segment displayed, proceed to that segment and continue
troubleshooting.
m. SEGMENT 11
1. VDT DISPLAY
TEMP CONTROLLER SOURCE B IN ERROR TEMPERATURE READS #.## AND SHOULD READ
35.00.
PERFORM THE FOLLOWING REPLACE ACTIONS IN THE ORDER LISTED.
(REFERENCE TM 11-6625-3081-23 FOR REPLACEMENT PROCEDURES.)
REF. DES.
PART NO.
NAME
1. 1A1
13082765
CONTROL UNIT ASSY.
2. 2A3A1A1A2
13082972
SOURCE B
20 CORRECTIVE ACTION - SEGMENT 11
(a) Remove Control Unit Assembly 1A1.
Check contacts of mating connectors
and repair if required.
Install replacement 1A1 and perform AUGTEM
test.
(It is not mandatory to perform hand held temperature probe
portion of test.) If AUGTEM test passes, no further action required.
If this segment displayed again, proceed to (b). If another segment
displayed, proceed to that segment and continue troubleshooting.
(b) Remove Source B 2A3A1A1A2.
Check contacts of mating connectors and
repair if required.
Install replacement 2A3A1A1A2 and perform AUGTEM
test.
(It is not mandatory to perform hand held temperature probe
portion of test.) If AUGTEM test passes, no further action required.
If this segment displayed again, proceed to (c). If another segment
displayed, proceed to that segment and continue troubleshooting.
.
Change 2
7-203