TM 11-6625-3081-23
8-3. ANALOG TO DIGITAL CONVERTER ASSEM8LY 2A2A13 (cont)
8 - 3
73.
74.
75.
76.
77.
78.
If necessary, adjust R401 (31) until
front panel DATA DISPLAY indicates
as close to 0 as possible.
(Upper
limit is 4 and lower limit is
177774.)
Remove shorting plug from P4 of rf
cable assembly. Connect P4 of rf
cable assembly to unconnected
connector of 13081706-69 cable.
If necessary, adjust R405 (32) until
front panel DATA DISPLAY indicates
as close to 60000 as possible.
(Upper limit is 60004 and lower
limit is 57774.)
Disconnect P4 of rf cable assembly
from 13081706-69 cable.
Press front panel POWER switch to
remove power.
CAUTION
The A/D converter contains parts and
assemblies sensitive to electro-
static discharge (ESD). Use ESD
precautionary procedures (para 2-21)
when touching, removing, or
inserting CCAs.
Remove rf cable assembly from CCA.
Remove CCA extender from slot J5.
Remove CCA from CCA extender.
Install CCA in slot J5.
8-12
Change 8