TM 11-6625-3081-23
8-3. ANALOG TO DIGITAL CONVERTER ASSEMBLY 2A2A13 (cont)
8 -3
61. If necessary, adjust R201 (27) until
front panel DATA DISPLAY indicates
as close to O as possible.
(Upper
limit is 4 and lower limit is
177774.)
62. Remove shorting plug from P2 of rf
cable assembly.
Connect P2 of rf
cable assembly to unconnected
connector of 13081706-69 cable.
63. If necessary, adjust R205 (28) until
front panel DATA DISPLAY indicates
as close to 60000 as possible.
(Upper limit is 60004 and lower
limit is 57774.)
64. Disconnect P2 of rf cable assembly
from 13081706-69 cable.
65. Set front panel digiswitches to
100220.
Install shorting plug on P3
of rf cable assembly.
66. Press front panel ANALOG switch.
67. If necessary, adjust R301 (29) until
front panel DATA DISPLAY indicates
as close to O as possible.
(Upper
limit is 4 and lower limit is
177774.)
68. Remove shorting plug from P3 of rf
cable assembly.
Connect P3 of rf
cable assembly to unconnected con-
nector of 13081706-69 cable.
69. If necessary, adjust R305 (30) until
front panel DATA DISPLAY indicates
as close to 60000 as possible.
(Upper limit is 60004 and lower
limit is 57774.)
70. Disconnect P3 of rf cable assembly
from 13081706-69 cable.
71. Set front panel digiswitches to
100230.
Install shorting plug on P4
of rf cable assembly.
72. Press front panel ANALOG switch.
Change 9
8-11