TM 11-6625-3081-23
2-14. MODULE DESCRIPTIONS (cont)
2-14
e. COMMON MODULES (cont)
Input Code
VIDEOST
EOBST
2-88
Change 9
Description
This module tests the portable video monitor. The portable
monitor is first checked for uniform brightness.
The operator
is prompted to observe the portable monitor for uniform bright-
ness over the entire screen, and to press yes if the brightness
is acceptable and no if it is unacceptable.
If unacceptable,
the operator is prompted to replace portable monitor.
The operator is prompted to place the FULL UNDER switch to the
under position and adjust the HEIGHT control until the vertical
and horizontal borders are about the same on all four sides.
The operator is prompted to press yes if acceptable and no
otherwise.
If no, the operator is prompted to replace the
portable monitor.
The operator is next prompted to place the FULL UNDER switch to
the full position and verify that the display is now completely
illuminated and the border has been completely removed. The
operator presses yes if the display is completely illuminated,
no otherwise.
If no, the operator is prompted to replace the
portable monitor.
The operator is next prompted to observe that a fine vertical
bar pattern is visible on the monitor and to press yes if the
image is acceptable and no if the image is unacceptable.
If no,
the operator is prompted to replace the portable monitor.
This module is an executive controller for the E/O augmentation
bench self-test.
This program is menu driven and controls all
input, output, and file handling. All other modules are called
by the executive controller as procedures. If a failure is
encountered during a specific module the failure will be
recorded and the operator will be given the option to continue
or exit.
This module contains 17 utility programs used in conjunction
with several items of EO bench equipment.
It is menu driven.
The following are covered:
MULTIPROGRAMMER DIG IN/OUT - Program a digital output CCA or
read a digital input CCA.
READ TEMPERATURE SENSORS - Read the following temperatures:
Mirror A transmission, mirror B transmission, source A,
source B, source C, OSA PMT, and temperature probe.
MAINT