TM 11-6625-3081-23
2-17. OPERATOR INSTRUCTION MESSAGES (cont)
2-17
(22) 1.
CONNECT P13 OF CABLE
13083701 TO PROGRAMMABLE DIGITIZER A2A1
CHANNEL A INPUT A (26).
2. CONNECT CURRENT PROBE AMPLIFIER
TYPE 134 TO PROGRAMMABLE DIGITIZER A2A1
CHANNEL A INPUT B (27).
3. SET CURRENT/DIV KNOB OF CURRENT PROBE
AMPLIFIER TO 1A.
4. SET DEFLECTION FACTOR SWITCH OF
CURRENT PROBE AMPLIFIER TO P6021/O.4MV.
5. CONNECT BNC CONNECTOR OF AC CURRENT
PROBE P6021 TO CURRENT PROBE AMPLIFIER.
PRESS PROCEED FOR ADDITIONAL
INSTRUCTIONS.
6. CONNECT DC POWER INPUT OF CURRENT
PROBE AMPLIFIER TO DC POWER OUTPUT OF
POWER SUPPLY FOR TYPE 134 (015-0058-01).
7.
CONNECT AC POWER CORD PLUG OF POWER
SUPPLY INTO E/O BENCH A2A4J3 (29).
8. CONNECT P1 OF CABLE 13081706 TO
PROGRAMMABLE DIGITIZER CHANNEL B INPUT A
(22) AND P2 TO EOB ELECTRONICS DRAWER J22
(30).
9.
CONNECT CURRENT PROBE TIP TO LOOP IN
SHORTING PLUG 13083962-559. NOTE:
CURRENT PROBE TIP ARROW MUST POINT TO
SAME DIRECTION AS SHORTING PLUG ARROW.
PRESS PROCEED WHEN COMPLETE.
Change 3
2-128.3