TM 11-6625-3081-23
2-17. OPERATOR INSTRUCTION MESSAGES (cont)
2-17
(5) 1.
CONNECT P1 OF CABLE 13083701
TO MAJOR TEST ADAPTER J2 (11).
2.
CONNECT P2 OF CABLE 13083701 TO MAJOR
TEST ADAPTER J3 (12).
PRESS PROCEED WHEN COMPLETE.
(6) 1. OPEN OSA ASSEMBLY A1A1 DOOR
(13 ) IF NOT ALREADY OPEN.
2. DISCONNECT P2 OF OSA CABLE
13081941-19 FROM OSA TEE CONNECTOR
A1A1J14 (14).
NOTE:
TEE CONNECTOR IS LOCATED IN
LOWER-LEFT FRONT CORNER.
HIGH VOLTAGE
LETHAL SHOCK HAZARD OR EQUIPMENT
DAMAGE.
A 16,000 VOLT. POTENTIAL IS
PRESENT ON THE W13 CABLE WHEN SYSTEM
IS IN OPERATION.
CABLE W13P1 MUST
ALWAYS BE CONNECTED TO CONTROL PANEL
ASSEMBLY CONNECTOR J1 OR TERMINATED
AT THE OSA HIGH VOLTAGE TERMINATION
CONNECTOR J1.
3. CONNECT CABLE ASSEMBLY (P/N 13081942)
2A1A1W13-P1 TO J1 OF OSA HIGH VOLTAGE
TERMINATION CONNECTOR (15).
4.
CLOSE OSA ASSEMBLY A1A1 DOOR (13).
PRESS PROCEED WHEN COMPLETE.
2-126
Change 3